Medgas100number2

Reviewed by Editorial Team
The ProProfs editorial team is comprised of experienced subject matter experts. They've collectively created over 10,000 quizzes and lessons, serving over 100 million users. Our team includes in-house content moderators and subject matter experts, as well as a global network of rigorously trained contributors. All adhere to our comprehensive editorial guidelines, ensuring the delivery of high-quality content.
Learn about Our Editorial Process
| By Pbarton
P
Pbarton
Community Contributor
Quizzes Created: 4 | Total Attempts: 22,717
| Attempts: 1,776
SettingsSettings
Please wait...
  • 1/100 Questions

    Station inlets for WAGD service shall be legibly marked to identify them and shall be of a type appropriate for the flow and vacuum level required by the facility’s gas anesthetic machines.

    • True
    • False
Please wait...
About This Quiz

The MEDgas100number2 quiz assesses knowledge on installation and maintenance of medical gas systems. Topics include pipe joint brazing, use of purge gases, and alarm system thresholds, crucial for professionals in healthcare facility management and medical system engineering.

Medgas100number2 - Quiz

Quiz Preview

  • 2. 

    Where instrument air systems are provided with standby headers, in the event of a compressor failure, the standby header shall automatically serve the system.

    • True

    • False

    Correct Answer
    A. True
    Explanation
    In instrument air systems where standby headers are present, if a compressor fails, the standby header will automatically take over and supply the system. This means that even if one compressor fails, the system will still have a backup source of air to ensure uninterrupted operation. Therefore, the statement is true.

    Rate this question:

  • 3. 

    Locations for central supply systems and the storage of medical gases shall be secured with lockable doors or gates.

    • True

    • False

    Correct Answer
    A. True
    Explanation
    The correct answer is true because it is important to secure the locations for central supply systems and the storage of medical gases with lockable doors or gates. This helps to prevent unauthorized access and ensures the safety and security of the medical gases, which are crucial for patient care. By having lockable doors or gates, healthcare facilities can control access to these areas and reduce the risk of theft, tampering, or misuse of the medical gases.

    Rate this question:

  • 4. 

    The presence and correctness of labeling shall be verified.

    • True

    • False

    Correct Answer
    A. True
    Explanation
    Reference: 5.1.12.3.13

    Rate this question:

  • 5. 

    While brazing medical-gas copper joints, a purge gas shall flow continuously to prevent the formation of copper oxide on the inside surface of the joint. What shall be used as a purge gas?

    • Oil-free, dry Argon MD

    • Oil-free, dry Helium NF

    • Oil-free, dry Nitrogen NF

    • Any inert gas

    Correct Answer
    A. Oil-free, dry Nitrogen NF
    Explanation
    During the brazing process of medical-gas copper joints, a purge gas is required to prevent the formation of copper oxide on the inside surface of the joint. This is because copper oxide can contaminate the joint and compromise its integrity. The ideal purge gas for this purpose should be oil-free and dry to avoid introducing any impurities into the joint. Nitrogen NF (National Formulary) is a high-purity nitrogen gas that meets the required standards for medical applications. It is inert, meaning it does not react with copper or other materials, making it suitable for purging and ensuring a clean joint.

    Rate this question:

  • 6. 

    Compressors for medical air shall be designed to prevent the introduction of contaminants or liquid into the piping system.

    • True

    • False

    Correct Answer
    A. True
    Explanation
    Compressors for medical air need to be designed in a way that ensures contaminants or liquid cannot enter the piping system. This is crucial for maintaining the purity and safety of the medical air supply, as any contaminants or liquid could potentially harm patients or compromise the effectiveness of medical procedures. Therefore, it is important to have compressors specifically designed to prevent the introduction of such substances into the piping system.

    Rate this question:

  • 7. 

    While brazing medical-gas copper joints, a purge gas shall flow continuously to prevent the formation of copper oxide on the inside surface of the joint. What shall be used as a purge gas?

    • Oil-free, dry Argon MD

    • Oil-free, dry Helium NF

    • Oil-free, dry Nitrogen NF

    • Any inert gas

    Correct Answer
    A. Oil-free, dry Nitrogen NF
    Explanation
    Reference: 5.1.10.5.5.1

    Rate this question:

  • 8. 

    Each station outlet/inlet for medical gas or vacuum shall be ____________.

    • Interchangeable

    • Gas-specific

    • Threaded

    • Quick-coupler

    Correct Answer
    A. Gas-specific
    Explanation
    Each station outlet/inlet for medical gas or vacuum must be gas-specific, meaning that it is designed and designated for a specific type of gas. This ensures that the correct gas is being delivered to the intended location and reduces the risk of accidental mix-ups or cross-contamination. Gas-specific outlets/inlets help maintain the safety and effectiveness of medical gas systems in healthcare facilities.

    Rate this question:

  • 9. 

    Receivers for medical air shall be equipped with _____________.

    • Relief valve

    • Automatic and manual drains

    • Sight glass

    • All of the above

    Correct Answer
    A. All of the above
    Explanation
    Receivers for medical air need to be equipped with a relief valve to ensure that the pressure does not exceed safe limits. They also need to have automatic and manual drains to remove any condensate or contaminants that may accumulate in the system. Additionally, a sight glass is necessary to allow visual inspection of the air quality. Therefore, all of the above options are required for receivers of medical air.

    Rate this question:

  • 10. 

    Relief valves shall be vented to the ___________, except compressed air systems having less than 3000 cubic feet at STP.

    • Floor

    • Sewer drain

    • Back into the system

    • Outside

    Correct Answer
    A. Outside
    Explanation
    Relief valves are designed to release excess pressure in a system to prevent damage or failure. To ensure the safe operation of the relief valve, it is important to vent the released pressure outside the system. This allows the pressure to be safely discharged into the atmosphere, away from any potential hazards or sensitive areas. However, for compressed air systems with a small volume (less than 3000 cubic feet at standard temperature and pressure), venting to the outside may not be necessary as the amount of pressure released is minimal.

    Rate this question:

  • 11. 

    All pressure relief valves shall be made of brass, bronze or ____________.

    • Stainless steel

    • Carbon steel

    • Cast iron

    • Copper

    Correct Answer
    A. Stainless steel
    Explanation
    Pressure relief valves are an essential component in various systems to prevent overpressure and ensure safety. These valves are subjected to high-pressure conditions, and therefore, they need to be made of materials that can withstand such conditions without corroding or deforming. Brass, bronze, and stainless steel are commonly used materials for pressure relief valves due to their high strength, corrosion resistance, and durability. Carbon steel, cast iron, and copper are not suitable choices for pressure relief valves as they may not possess the required properties to withstand high-pressure conditions. Therefore, stainless steel is the correct answer for the given question.

    Rate this question:

  • 12. 

    On system renovations where no piping is changed, functional testing shall be performed to assure no other equipment in the system has been adversely impacted.

    • True

    • False

    Correct Answer
    A. True
    Explanation
    Reference: 5.1.12.1.12.1

    Rate this question:

  • 13. 

    Hangers for copper tube shall be plastic-coated, copper finishes or otherwise insulated from the tube

    • True

    • False

    Correct Answer
    A. True
    Explanation
    Reference: 5.1.10.10.4.3 (.4)

    Rate this question:

  • 14. 

    Which of the following filler metals shall be used to braze copper-to-copper joints in medical gas/vacuum distribution piping?

    • BAg series

    • BAu series

    • BCuP series

    • BNi series

    Correct Answer
    A. BCuP series
    Explanation
    The BCuP series filler metals should be used to braze copper-to-copper joints in medical gas/vacuum distribution piping.

    Rate this question:

  • 15. 

    Tubing for positive pressure medical gas systems delivered from the manufacturer shall be__________________.

    • Plugged or capped

    • Harddrawn seamless

    • Identified as “oxy” or “med”

    • All of the above

    Correct Answer
    A. All of the above
    Explanation
    The tubing for positive pressure medical gas systems delivered from the manufacturer should be plugged or capped to prevent any contamination or leakage. It should also be harddrawn seamless to ensure durability and prevent any cracks or leaks. Additionally, the tubing should be identified as "oxy" or "med" to clearly indicate its purpose and prevent any mix-ups or confusion. Therefore, all of the above options are correct requirements for the tubing of positive pressure medical gas systems.

    Rate this question:

  • 16. 

    All valves except zone valve box assemblies shall be ________________.

    • In a secured area

    • Locked or latched

    • Labeled appropriately

    • All of the above

    Correct Answer
    A. All of the above
    Explanation
    All valves except zone valve box assemblies shall be secured in a secured area, locked or latched, and labeled appropriately. This ensures that the valves are kept in a safe and controlled environment, preventing unauthorized access or tampering. Additionally, labeling the valves appropriately allows for easy identification and maintenance. By implementing all of these measures, the valves can be effectively managed and maintained, ensuring the safety and functionality of the system.

    Rate this question:

  • 17. 

    In a Level 3 vacuum system installed with plastic piping, the pipe shall be polyvinyl chloride (PVC) plastic with a minimum wall thickness of ____________.

    • Schedule 10

    • Schedule 20

    • Schedule 40

    • Any of the above

    Correct Answer
    A. Schedule 40
    Explanation
    Reference: 5.3.10.2.3
    ANSWERS

    Rate this question:

  • 18. 

    When using the autogenous orbital welding process, joints shall be continuously purged inside and outside with an inert gas as qualified with the procedure.

    • True

    • False

    Correct Answer
    A. True
    Explanation
    Reference: 5.1.10.5.5.12

    Rate this question:

  • 19. 

    A signal from each local alarm system shall be connected to the master alarm panels to indicate that a problem is present with the source equipment.

    • True

    • False

    Correct Answer
    A. True
    Explanation
    Reference: 5.1.9.5.2

    Rate this question:

  • 20. 

    All medical gas alarm systems shall have a ____________ audible indication for each condition monitored.

    • Adjustable

    • Cancelable

    • Optional

    • All of the above

    Correct Answer
    A. Cancelable
    Explanation
    Medical gas alarm systems are required to have a cancelable audible indication for each condition monitored. This means that the alarm can be silenced or canceled once the condition being monitored has been resolved or addressed. This feature allows healthcare professionals to acknowledge and respond to the alarm, reducing unnecessary noise and preventing alarm fatigue.

    Rate this question:

  • 21. 

    The factory installed tubing on station outlets/inlets shall extend no further than ______ from the body of the terminal.

    • 6"

    • 8"

    • 10"

    • 12"

    Correct Answer
    A. 8"
    Explanation
    The factory installed tubing on station outlets/inlets should not extend more than 8 inches from the body of the terminal. This is to ensure that the tubing does not become too long and cause any potential hazards or obstructions. By limiting the length to 8 inches, it allows for proper installation and functionality of the terminal without any unnecessary risks.

    Rate this question:

  • 22. 

    Installers of bulk cryogenic liquid systems shall be qualified to meet CGA M-1 Standard, Guide for Medical Gas Installations at Consumers Sites.

    • True

    • False

    Correct Answer
    A. True
    Explanation
    The statement is true because according to the given information, installers of bulk cryogenic liquid systems need to be qualified to meet CGA M-1 Standard, which is the Guide for Medical Gas Installations at Consumer Sites. This standard ensures that the installation of these systems is done correctly and safely in medical facilities where cryogenic liquids are used.

    Rate this question:

  • 23. 

    Cylinders in use and in storage shall be prevented from reaching temperatures in excess of _______________.

    • 100º F

    • 130º F

    • 160º F

    • 190º F

    Correct Answer
    A. 130º F
    Explanation
    Cylinders in use and in storage should be prevented from reaching temperatures in excess of 130º F. This is important because exceeding this temperature can lead to the expansion of gases inside the cylinder, which can cause the cylinder to rupture or explode. It is necessary to maintain a safe temperature range to ensure the integrity and stability of the cylinders and to prevent any potential hazards.

    Rate this question:

  • 24. 

    For the initial pressure test, all brazed joints shall be examined for leakage by means of soapy water or other equally effective means that are safe for use with oxygen.

    • True

    • False

    Correct Answer
    A. True
    Explanation
    Reference: 5.1.12.2.3.6

    Rate this question:

  • 25. 

    Which of the following filler metals shall be used to braze copper-to-copper joints in medical gas/vacuum distribution piping?

    • BAg series

    • BAu series

    • BCuP series

    • BNi series

    Correct Answer
    A. BCuP series
    Explanation
    Reference: 5.1.10.5.1.5

    Rate this question:

  • 26. 

    Computer systems used as a master alarm shall be continuously attended or shall provide remote signaling of responsible parties through _________. .

    • Pagers

    • Telephone

    • Cell-phone

    • Any of the above

    Correct Answer
    A. Any of the above
    Explanation
    Reference: 5.1.9.4.1 (2)

    Rate this question:

  • 27. 

    The master alarm system shall activate when the pressure in the main supply lines increase or decrease by __________ percent.

    • 10

    • 20

    • 40

    • 50

    Correct Answer
    A. 20
    Explanation
    The master alarm system is designed to activate when there is a significant change in pressure in the main supply lines. In this case, the correct answer is 20 percent, which means that the alarm will be triggered if the pressure increases or decreases by 20 percent. This threshold ensures that any significant deviations from the normal pressure levels are detected and alerted, allowing for prompt action to be taken to address any potential issues with the supply lines.

    Rate this question:

  • 28. 

    __________ connectors shall connect the air compressors with their intake and outlet piping.

    • Anti-vibration

    • Dielectric

    • Insulated

    • Flexible

    Correct Answer
    A. Flexible
    Explanation
    Flexible connectors are used to connect the air compressors with their intake and outlet piping. These connectors are able to absorb vibrations and movements, providing flexibility and preventing damage to the piping system. They allow for easy installation and maintenance, as well as reducing noise and vibration transmission. Therefore, flexible connectors are the most suitable option for connecting air compressors with their intake and outlet piping.

    Rate this question:

  • 29. 

    Where the Level 3 central supply is remote from the medical gas system use points, the main supply line shall be provided with an emergency shutoff valve so located in the single treatment facility as to be accessible from the use-point location.

    • True

    • False

    Correct Answer
    A. True
    Explanation
    Reference: 5.3.4.1.2

    Rate this question:

  • 30. 

    Area alarms shall be tested to verify the warning signals for all vital life support and critical care areas if the pressure in the piping system increases or decreases by ____________.

    • 60%

    • 40%

    • 20%

    • 10%

    Correct Answer
    A. 20%
    Explanation
    Reference: 5.1.12.3.5.3

    Rate this question:

  • 31. 

    Area alarm panels for medical gas systems shall indicate when line pressure in the area being monitored increases or decreases by ___________ percent.

    • 10

    • 15

    • 20

    • 25

    Correct Answer
    A. 20
    Explanation
    Reference: 5.1.9.3.2

    Rate this question:

  • 32. 

    The quality of the instrument air shall be _________________.

    • Filtered to .01 micron

    • Free of liquids

    • Free of hydrocarbon vapors

    • All of the above

    Correct Answer
    A. All of the above
    Explanation
    The correct answer is "All of the above." This means that the quality of the instrument air should meet all the mentioned criteria, which includes being filtered to .01 micron, free of liquids, and free of hydrocarbon vapors. These requirements ensure that the instrument air is clean and suitable for use in various instruments and systems.

    Rate this question:

  • 33. 

    Medical air sources shall be used only for _________ and calibration of medical respiratory devices.

    • Human respiration

    • Air-operating devices

    • Instrument air

    • All of the above

    Correct Answer
    A. Human respiration
    Explanation
    Medical air sources are specifically designed and regulated for the purpose of providing clean and safe air for human respiration. Therefore, they should only be used for this purpose and not for operating air-operating devices or for instrument air. Using medical air sources for anything other than human respiration could compromise the quality and safety of the air supply, which is crucial for patients relying on respiratory devices.

    Rate this question:

  • 34. 

    Louvered openings for natural ventilation shall have a minimum free area of __________.

    • 60 sq. in.

    • 62 sq. in.

    • 70 sq. in.

    • 72 sq. in.

    Correct Answer
    A. 72 sq. in.
    Explanation
    Louvered openings for natural ventilation should have a minimum free area of 72 sq. in. This means that the total area of the openings should be at least 72 square inches to allow for sufficient airflow. Having a larger free area ensures better ventilation and allows for more air to enter or exit a space, improving air quality and reducing the risk of stagnant air.

    Rate this question:

  • 35. 

    The verifier is required to test the proper operation of the valves installed in each medical gas andvacuum piping system. The verifier is also required to make a list of the rooms or areas controlled by each valve for each gas.

    • True

    • False

    Correct Answer
    A. True
    Explanation
    Reference: 5.1.12.3.4 (.1)

    Rate this question:

  • 36. 

    Runouts from horizontal mains or branches shall be taken off ________ the centerline of the pipeline.

    • Above

    • Below

    • At

    • Any of the above

    Correct Answer
    A. Above
    Explanation
    Runouts from horizontal mains or branches should be taken off above the centerline of the pipeline. This is because locating the runouts above the centerline ensures that any air or gas pockets in the pipeline can be easily released. Additionally, taking the runouts off above the centerline helps to prevent the accumulation of sediment or debris in the runouts, which could potentially obstruct the flow of fluid in the pipeline.

    Rate this question:

  • 37. 

    Carbon monoxide shall activate the required alarms when the CO level exceeds ________.

    • 5 ppm

    • 10 ppm

    • 15 ppm

    • 20 ppm

    Correct Answer
    A. 10 ppm
    Explanation
    Carbon monoxide shall activate the required alarms when the CO level exceeds 10 ppm. This means that the alarms will only be triggered when the concentration of carbon monoxide in the air reaches or exceeds 10 parts per million. This is an important safety measure as carbon monoxide is a highly toxic gas that can be lethal in high concentrations. By setting the activation threshold at 10 ppm, the alarms will alert individuals to the presence of carbon monoxide before it reaches dangerous levels, allowing them to take immediate action to protect themselves and evacuate if necessary.

    Rate this question:

  • 38. 

    Where charcoal filters are required, a colormetric hydrocarbon indicator shall be provided.

    • True

    • False

    Correct Answer
    A. True
    Explanation
    In situations where charcoal filters are necessary, it is important to have a colormetric hydrocarbon indicator. This indicator helps to detect the presence of hydrocarbons in the air or water that is being filtered. By providing this indicator, it ensures that the filters are effectively removing any harmful hydrocarbons and maintaining the quality of the filtered substance. Therefore, the statement is true.

    Rate this question:

  • 39. 

    Level 1 medical air compressors shall be sized to serve the peak calculated demand and in no case shall there be fewer than _____ compressors.

    • One

    • Two

    • Three

    • Four

    Correct Answer
    A. Two
    Explanation
    Level 1 medical air compressors need to be sized to meet the peak calculated demand. This ensures that there is enough compressed air available for medical procedures. Additionally, there should be a minimum of two compressors to ensure redundancy and prevent any disruptions in the supply of medical air. Having multiple compressors also allows for maintenance and repairs to be carried out on one compressor while the other one continues to provide the necessary air supply.

    Rate this question:

  • 40. 

    Manifolds for gas cylinders without reserve shall consist of ___________ equal headers; each with an average day’s supply, but not fewer than _________ connections.

    • Four – four

    • Three – three

    • Two – two

    • One – one

    Correct Answer
    A. Two – two
    Explanation
    Manifolds for gas cylinders without reserve should consist of two equal headers, each with an average day's supply, but not fewer than two connections. This means that there should be two headers, each capable of supplying enough gas for an average day's usage, and each header should have at least two connections. This ensures that there is redundancy in the system, with two headers providing backup in case one fails, and each header having multiple connections to accommodate the required number of cylinders.

    Rate this question:

  • 41. 

    Liquid containers shall have additional product identification visible from all directions with a minimum of _____ inch high letters such as a 360 degree wrap-a-round tape for medical liquid containers.

    • 1

    • 2

    • 3

    • 4

    Correct Answer
    A. 2
    Explanation
    Liquid containers should have additional product identification visible from all directions with a minimum of 2-inch high letters. This is important for ensuring clear and easy identification of the product, especially in medical settings where accuracy and efficiency are crucial. Having the identification visible from all directions helps to prevent any confusion or mistakes in handling or administering the liquid.

    Rate this question:

  • 42. 

    Since dew point and carbon monoxide monitors are located upstream of the source valve, demand check fittings are not required.

    • True

    • False

    Correct Answer
    A. False
    Explanation
    The statement is false because demand check fittings are required for dew point and carbon monoxide monitors, even if they are located upstream of the source valve. Demand check fittings are necessary to ensure proper functioning and accuracy of these monitors.

    Rate this question:

  • 43. 

    Where shall pressure indicators be provided for medical gas systems?

    • On the main supply line

    • At the area alarm

    • Outlet/inlet side of zone valve

    • All of the above

    Correct Answer
    A. All of the above
    Explanation
    Pressure indicators should be provided for medical gas systems at all of the mentioned locations. This is because pressure indicators are essential for monitoring and maintaining the correct pressure levels in the system. By having pressure indicators on the main supply line, at the area alarm, and on the outlet/inlet side of zone valves, healthcare professionals can easily monitor the pressure levels at different points in the system and ensure that the medical gas supply is safe and reliable.

    Rate this question:

  • 44. 

    The scale range of positive pressure analog indicators shall have the normal reading fall within the_________ percentage of the gage scale.

    • 25

    • 50

    • 75

    • 100

    Correct Answer
    A. 50
    Explanation
    The scale range of positive pressure analog indicators should have the normal reading fall within 50% of the gage scale. This means that the midpoint of the scale should represent the normal operating pressure. Having the normal reading fall within 50% of the scale allows for a sufficient range to measure higher or lower pressures without going beyond the limits of the indicator.

    Rate this question:

  • 45. 

    A Level 1 dedicated WAGD source shall consist of _____ or more producers, each shall be sufficient to serve the peak calculated demand.

    • One

    • Two

    • Three

    • All of the above

    Correct Answer
    A. Two
    Explanation
    A Level 1 dedicated WAGD source shall consist of two or more producers, each sufficient to serve the peak calculated demand. This means that in order to meet the demand for WAGD, there should be at least two producers available. Having only one producer may not be enough to meet the peak demand. Therefore, the correct answer is Two.

    Rate this question:

  • 46. 

    The compressor air intakes shall be located outdoors above roof level at a minimum distance of__________ above the ground.

    • 40 ft.

    • 30 ft.

    • 20 ft.

    • 10 ft.

    Correct Answer
    A. 20 ft.
    Explanation
    The compressor air intakes should be located outdoors above roof level to ensure that they are not affected by any obstructions or contaminants present at ground level. Placing them at a minimum distance of 20 ft above the ground helps to prevent the intake of dust, debris, and other pollutants that could potentially damage the compressor and affect its performance. This height also allows for proper air circulation and reduces the risk of any potential hazards.

    Rate this question:

  • 47. 

    Bulk cryogenic liquid system sites shall have allowance for at least ____ of clearance around the storage container, vaporizer and the pressure-regulating manifold.

    • 2 feet

    • 3 feet

    • 4 feet

    • 5 feet

    Correct Answer
    A. 3 feet
    Explanation
    Bulk cryogenic liquid systems require a certain amount of clearance around the storage container, vaporizer, and pressure-regulating manifold for safety purposes. The correct answer is 3 feet. This clearance allows for adequate ventilation and prevents any potential hazards or accidents that may occur due to heat buildup or pressure fluctuations. Having a sufficient distance between these components ensures the safe operation of the cryogenic liquid system.

    Rate this question:

  • 48. 

    Which group or agency shall review the inspection and testing records prior to the use of all systems to assure that all findings and results of the inspection and testing have been successfully completed?

    • Responsible facility authority

    • Maintenance supervisor

    • Head nurse

    • Any of the above

    Correct Answer
    A. Responsible facility authority
    Explanation
    Reference: 5.1.12.1.8

    Rate this question:

  • 49. 

    A shutoff valve shall be placed at the immediate connection of each source system to the distribution piping system. What is the name given to the valve?

    • Main

    • Supply

    • Source

    • Isolation

    Correct Answer
    A. Source
    Explanation
    The valve referred to in the question is called the "source" valve. This valve is placed at the immediate connection of each source system to the distribution piping system. It is responsible for shutting off the flow of fluid or gas from the source system into the distribution system.

    Rate this question:

Quiz Review Timeline (Updated): Mar 21, 2023 +

Our quizzes are rigorously reviewed, monitored and continuously updated by our expert board to maintain accuracy, relevance, and timeliness.

  • Current Version
  • Mar 21, 2023
    Quiz Edited by
    ProProfs Editorial Team
  • Nov 08, 2010
    Quiz Created by
    Pbarton
Back to Top Back to top
Advertisement
×

Wait!
Here's an interesting quiz for you.

We have other quizzes matching your interest.